ATL Turbidity Probe
Brand: KOBOLD Instruments, IncATL-F
The Kobold Turbidity Probe ATL-F is a high-precision, single channel absorption probe. The probe uses the light in the visible (VIS) range at a selected wavelength of 430 nm. An optical filter on the lamp side adapts the wavelength for specific applications and allows colour measurements with great accuracy (e.g. phase separation beer/water) at 430 nm.
ATL-N
The Kobold Turbidity Probe ATL-N is a high-precision NIR single channel absorption probe. The ATL-N uses light in the Near Infrared (NIR) from 730 to 970 nm. The optical filter on the lamp side adapts the wavelength for specific applications and allows colour-insensitive concentration measurements in the near infrared range. The ATL-N probe comes with two different optical path lengths (OPL) for process versatility.
Specifications
Brand
Alarm
- 2 Independent Adjustable SPDT Contacts
Ambient Temperature Range
- -20° to 70° C (-4° to 158° F)
Applications
- Phase Separation
Approvals & Certifications
- CE
- ISO 9001: 2000
- PED
BSP Connection
- G 1-1/4”
Cable Length
- 164 feet (50 metres)
Consumption
- 30 VA
Display
- 3-Digit LED Display
Electrical Connectors
- Plug Connection
Enclosure Finish
- Electro-Polishing Ra < 0.8 um (Standard)
Enclosure, Body Material
- 316Ti Stainless Steel
Environmental Protection
- IP20 - Rear
- IP40 - Front
- IP65
Gasket Material
- Chemraz
- EPDM
- Kalrez 6375
- Silicone
Length
- 30 mm - Nominal
- 60 mm - Nominal
- Insertion Depth, Max.
Linearity
- < ± 2 % of Respective Measuring Range
Load
- 500 ohms, Max.
Materials of Construction
- 316L Stainless Steel - Adapter
Metric Connection
- M5 Female
Mounting
- Rack-Mounted
Operating Temperature
- 0° to 50° C (32° to 122° F)
Output
- 0/4-20 mA (Galvanically Isolated)
Power Requirement
- 115 Vac
- 230 Vac
- 24 Vac
- 24 Vdc
Pressure Rating
- psi: 145 (kPa: 1,000, bar: 10)
Process Temperature
- 0° to 100° C (32° to 212° F)
Resolution
- < ± 0.5 % of Respective Measuring Range
Response Time
- 1 second
Size
- O-Ring Diameter: 18.64 x 3.53 mm (Port Gasket)
Wavelength
- 430 to 970 Nm
Wetted Materials
- 316L Stainless Steel
Window Material
- Sapphire (without Gasket)