Accuracy
- ≤ ± 0.05 % of Span - Including Non-Linearity, Hysteresis, Non-Repeatability, Zero Offset & End Value Deviation Per IEC 61298-2
- ≤ ± 0.1 % of Span
Ambient Temperature Range
- -20° to 80° C (-4° to 176° F)
Applications
- Calibration Technology
- Laboratories
- Measurement & Test Benches
- Plant Construction & Machine Building
Approvals & Certifications
- BelCT
- CE
- CPT
- CRN - Safety (e.g. Electr. Safety, Overpressure, to), Canada
- Directive 97/23/EC - PS > 200 bar, Module A, Pressure Accessory
- EAC
- EMC Directive - 2004/108/EC, EN 61326 Emission (Group 1, Class B) & Immunity (Industrial Application)
- GOST-R
- KazlnMetr
- RoHS Conformity (Not with Bayonet Connector)
- UkrSEPRO
- UZSTANDARD
BSP Connection
- G 1” B Male with Flush Diaphragm (EN 837)
- G 1/2” B Male (EN 837) / G 1/2” Male with G 1/4” Female / G 1/2” B Male with Flush Diaphragm (EN 837)
- G 1/4” - B Male (EN 837) / A Male (DIN 3852-E)
- G 1/4” Female (EN 837)
Consumption
- 40 mA - USB
- 60 mA - CANopen
- Maximum
Electrical Connectors
- Angular Connector DIN 175301-803 A
- Bayonet Connector
- Cable Outlet
- Circular Connector M12 x 1 (4-Pin) / M16 x 0.75 (5-Pin) / M12 x 1 (5-Pin), CANopen
- USB
Materials of Construction
- Copper
- FKM
- FPM
- NBR
- Stainless Steel
Metric Connection
- M18 x 1.5 Male with G 1/4” Female
Non-Linearity
- ≤ ± 0.04 % of Span BFSL (per IEC 61298-2)
NPT Connection
- 1/2” NPT Male (ANSI/ASME B1.20.1)
- 1/4” NPT Male (ANSI/ASME B1.20.1)
Output
- 0-10 Vdc (Voltage 3-Wire)
- 0-20 mA (Current 3-Wire)
- 0-5 Vdc (Voltage 3-Wire)
- 4-20 mA (Current 2-Wire / 3-Wire)
- CANopen - Per CiA DS 404
- USB - Per P-30/P-31 Interface Protocol
Overpressure & Static Pressure Limits
- psi: 20 to 15,000 (kPa: 150 to 150,000, bar: 1.5 to 1,500)
Power Supply
- 14-30 Vdc - Output Signal: 0-10 Vdc
- 4.5-5.5 Vdc - Output Signal: USB
- 9-30 Vdc - Output Signal: 4-20 mA (2-Wire/3-Wire), 0-20 mA, 3-Wire, 0-5 Vdc, CANopen
Pressure Range
- bar: 0 to 1,000
- kPa: 0 to 100,000
- psi: 0 to 10,000
Shock
- 200 G (IEC 60068-2-27, Mechanical)
Span & Zero Adjustments
- Span: -50 to 5 % of Span
- Zero Point: -5 to 10 % of Span
Storage Temperature
- -40° to 85° C (-40° to 185° F)
Temperature Range
- -20° to 105° C (-4° to 221° F)
Vibration
- 10 G (IEC 60068-2-6, Under Resonance)
Wetted Materials
- Elgiloy for Measuring Ranges > 25 bar
- Stainless Steel