Nominal Size
- MAG 3100: DN 15 ... DN 2000 (1/2” ... 78”)
- MAG 3100 HT: DN 15 … DN 300 (1/2” ... 12”)
Excitation Frequency (Mains Supply: 50 Hz/60 Hz)
- MAG 3100 - DN 80 - 150 (3” - 6”): 6.25 Hz/7.5 Hz
- MAG 3100 - DN 200 - 1200 (8” - 48”): 3.125 Hz/3.75 Hz
- MAG 3100 - DN 1400 - 2000 (54” - 78”): 1.5625 Hz/1.875 Hz
- MAG 3100 HT - DN 80 - 150 (3” - 6”): 6.25 Hz/7.5 Hz
- MAG 3100 HT - DN 200 - 300 (8” - 12”): 3.125 Hz/3.75 Hz
- MAG 3100 - DN 15 - 65 (1/2” - 2-1/2”): 12.5 Hz/15 Hz
- MAG 3100 HT - DN 15 - 65 (1/2” - 2-1/2”): 12.5 Hz/15 Hz
Flange
- EN 1092-1, Raised Face Flange
- ANSI B16.5 (~BS 1560), Raised Face Flange
- AS 2129, Raised Face Flange
- AS 4087, Raised Face Flange
- JIS B 2220:2004
Ambient Temperature
- Standard -40° to 100° C (-40° to +212° F)
- Ex Sensor: -20° to +60° C (-46° to +140° F)
MAG 3100 Compact with Transmitter
- MAG 5000/6000 -20 … +60° C (-4 … +140° F)
- MAG 6000 I -20 … +60° C (-4 … +140° F)
- MAG 6000 I Ex -20 … +60° C (-4 … +140° F)
MAG 3100 HT Compact with Transmitter
- MAG 5000/6000 -20 … +60° C (-4 … +140° F)
- MAG 6000 I -20 … +60° C (-4 … +140° F)
- MAG 6000 I Ex -20 … +60° C (-4 … +140° F)
MAG 3100 Operating Pressure
- Soft rubber 0.01 ... 100 bar (0.15 ... 1450 psi)
- EPDM 0.01 ... 40 bar (0.15 ... 580 psi)
- Linatex 0.01 ... 40 bar (0.15 ... 580 psi)
- Ebonite 0.01 ... 100 bar (0.15 ... 1450 psi)
- PTFE - DN ≤ 300 (≤ 12”): 0.3 ... 50 bar (4 ... 725 psi) - 350 ≤ DN ≤ 600 (14” ≤ DN ≤ 24”): 0.3 ... 40 bar (4 ... 580 psi)
- PFA - DN 15 ... 150 (1/2” … 6”): Vacuum 0.02 ... 50 bar (0.29 ... 725 psi)
MAG 3100 HT Operating Pressure
- PTFE Teflon - DN 15 ... 300 (1/2” … 12”): 0.3/0.6 ... 50 bar (4/8 ... 725 psi) (180° C (356° F)) Factory mounted grounding rings type E in stainless steel and stainless steel terminal box. Can only be used with remote transmitter.
- PFA - DN 15 ... 150 (1/2” … 6”): Vacuum 0.02 ... 50 bar (0.29 ... 725 psi)
Enclosure Rating
- IP67 to EN 60529/NEMA 4X/6, 1 mH₂O for 30 min
- Option: IP68 to EN 60529/NEMA 6P, 10 mH₂O cont.
Test Pressure
- 1.5 x PN (where applicable)
Mechanical Load (Vibration)
- 18 ... 1000 Hz random in x, y, z, directions for 2 hours according to EN 60068-2-36
- Sensor: 3.17 g rms
- Sensor with compact MAG 5000/6000 mounted transmitter: 3.17 g rms
- Sensor with compact MAG 6000 I/ 6000 I Ex mounted transmitter: 1.14 g rms
MAG 3100 Temperature of Medium
- Soft rubber 0 … +70° C (32 … 158° F)
- EPDM -10° to +70° C (14 … 158° F)
- Linatex (rubber) -40° to +70° C (-40° to +158° F) (for temperatures below -20° C (-4° F) AISI 304 or 316 flanges must be used)
- Ebonite 0° … 95° C (32 … 203° F)
- PTFE -20 … +100° C (-4 … +212° F)
- PFA -20 … +100° C (-4 … +212° F)
MAG 3100 HT Temperature of Medium
- 18 to 1000 Hz random in x, y, z, directions for 2 hours according to EN 60068-2-36
- Sensor: 3.17 g rms
- Sensor with compact MAG 5000/6000 mounted transmitter: 3.17 g rms
- Sensor with compact MAG 6000 I/ 6000 I Ex mounted transmitter: 1.14 g rms
- PTFE -20 … +150° C (-4 … +302° F)
- PTFE -20 … +180° C (-4° to +356° F) Factory mounted grounding rings type E in stainless steel and stainless steel terminal box. Can only be used with remote transmitter
- PFA -20 … +150° C (-4 … +300° F)
Flange and Housing Material
- Carbon steel ASTM A 105, with corrosion resistant coating
- Corrosivity category C4, according to ISO 12944-2 or stainless steel AISI 304/1.4301 flanges and carbon steel housing, with corrosion resistant coating
- Corrosivity category C4, according to ISO 12944-2 or stainless steel AISI 316L/1.4404 flanges and housing, polished
Electrode Material
- Stainless Steel AISI 316Ti/1.4571
- Hastelloy C-276/2.4819 (PFA: Hastelloy C-22/2.4602)
- Platinum/Iridium
- Titanium
- Tantalum
Grounding Electrode Material
- PTFE: no grounding electrodes
- PFA: optional in Hastelloy, Tantalum or Platinum
- MAG 3100: Soft rubber, EPDM, Linatex, Ebonite: available with grounding electrodes in stainless steel AISI 316Ti/1.4571 or Hastelloy
MAG 3100 Terminal Box (Remote Version Only)
- Standard fibre glass reinforced polyamide
- Option Stainless Steel AISI 316/1.4436
- Ex Stainless Steel AISI 316/1.4436
MAG 3100 HT Terminal Box Remote Version Only)
- Standard fibre glass reinforced polyamide (Max. 150° C (302° F))
- Stainless Steel AISI 316/1.4436
- Ex Stainless Steel AISI 316/1.4436
MAG 3100 Cable Entries
- Remote installation 2 x M20 or 2 x 1/2” NPT
- Compact installation - MAG 5000/MAG 6000: 4 x M20 or 4 x 1/2” NPT - MAG 6000 I: 2 x M25 or 2 x 1/2” NPT (for supply/output) - MAG 6000 I Ex: 2 x M25 or 2 x 1/2” NPT (for supply/output)
MAG 3100 HT Cable Entries
- Remote installation 2 x M20 or 2 x 1/2” NPT
Standard Production Calibration
- Zero-Point, 2 x 25 % and 2 x 90 %
MAG 3100 Special Calibration
- 5-Point calibration: 20 %, 40 %, 60 %, 80 %, 100 % of factory Qmax
- 10-Point calibration: ascending and descending at 20 %, 40 %, 60 %, 80 %, 100 % of factory Qmax
- Matched-pair calibration: default, 5-point or 10-point
Ex Sensor in Compact or Remote Version with MAG 6000 I Ex
- ATEX, FM, CSA, IECEx, EAC Ex, NEPSI - Zone 1 Ex d e ia IIC T6 Gb - Zone 1 Ex e ia IIC T6 Gb
- ATEX, FM, CSA, IECEx - Zone 21 Ex tD A21 IP67
- FM - XP IS Class I, Div. 1 Groups A, B, C, D - DIP Class II+III, Div. 1 Groups E, F, G
Standard Sensor With/Without MAG 5000/6000/6000 I
- FM - NI Class I, Div. 2 Groups A, B, C, D - NI Class I Zone 2 Groups IIC
MAG 3100 Drinking Water
- WRAS (WRc, BS690 cold water, GB)
- NSF/ANSI Standard 61 (Cold water, US)
- ACS listed (F)
- DVGW W270 (D)
- Belgaqua (B)
- MCERTS (GB) (EPDM or PTFE lining with AISI 316 or Hastelloy electrodes)
Pressure Equipment
- PED conforming: All EN 1092-1 flanges - 2014/68/EU
- CRN
Others
- EAC (Russia, Belarus, Kazakhstan)
- KCC (South Korea)
- CMC/CPA (China)